Product Details

Home / Products / LAMBDA 365+/465 UV/Vis Spectrophotometer

LAMBDA 365+/465 UV/Vis Spectrophotometer

LAMBDA 365+/465 UV/Vis Spectrophotometer

Model: LAMBDA 365+ / 465

The workhorse UV/Vis for universities and QA labs: 190 – 1100 nm. The 365+ uses a double-beam Czerny-Turner optical system with 0.5 – 20 nm variable bandwidth and 4 A absorbance; the 465 uses a photodiode array that captures a full spectrum in 20 ms. A large compartment takes over ten sampling accessories including autosamplers and thermostatted cell holders, with 21 CFR Part 11 compliant software.

Specifications

Models LAMBDA 365+ (double beam) / LAMBDA 465 (PDA)
Wavelength range 190 – 1100 nm
Spectral bandwidth 365+: 0.5 / 1 / 2 / 5 / 20 nm variable; 465: 1 nm
Optical system 365+: double-beam Czerny-Turner; 465: single-beam photodiode array
Photometric range up to 4 A
Wavelength accuracy ±0.1 nm at 656.1 nm; ±0.3 nm across range
Wavelength reproducibility better than 0.05 nm
Scan speed 365+: 7.5 – 12000 nm/min; 465: full spectrum in 20 ms
Stray light 465: < 0.03 %T at 340 nm
Light sources Pre-aligned deuterium + tungsten-halogen, automatic switching
Sample compartment Large compartment, more than 10 accessories supported
Software UV WinLab, 21 CFR Part 11 compliant
Dimensions 595 × 272 × 535 mm, 23.8 kg (365+)

Features

  • 190 – 1100 nm covers the full UV/Vis range for quantitation, kinetics and purity checks
  • 365+ double-beam Czerny-Turner design measures sample and reference together, so long kinetic runs do not drift
  • 0.5 – 20 nm variable bandwidth balances high resolution against throughput
  • 4 A photometric range takes concentrated samples without dilution
  • 365+ scans at 7.5 – 12000 nm/min to capture fast reaction kinetics
  • 465 photodiode array acquires a complete spectrum in 20 ms — ideal for dissolution and fast kinetics
  • Large compartment accepts multi-cell changers, thermostatted holders, autosamplers and more
  • UV WinLab ships with method templates and performance verification; the ES edition meets 21 CFR Part 11

Applications

  • Teaching: spectrophotometry fundamentals, Beer-Lambert verification, DNA and protein quantitation
  • Pharma and drug testing: dissolution, content uniformity, related substances, method development
  • Food and agriculture: total phenolics, flavonoids, nitrite, pigments and additives
  • Environmental monitoring: total phosphorus, total nitrogen, hexavalent chromium, COD
  • Life science: enzyme activity, protein concentration, nucleic acid purity (A260/A280)
  • Chemicals and materials: reaction kinetics, catalyst activity, in-line concentration monitoring
☎Call us ✉Enquiry
13838311486 Same number on WeChat