X-ray Diffractometer Specifications
1. X-ray tube and power
• Copper target as standard with K alpha at 1.5406 Angstrom, with cobalt, iron or chromium options
• Conventional power of 1.2 kilowatts or more, high resolution types at 1.6 kilowatts or more, microfocus types lower power but higher brightness
• Focal spot size and target material must be stated
• Water or air cooling chosen according to laboratory conditions
• Note: the target decides what can be measured, since iron and cobalt bearing samples fluoresce strongly under copper radiation
2. Goniometer
• Radius from 240 to 300 millimetres in theta theta or theta two theta geometry
• Two theta range of at least 3 to 150 degrees
• Angular reproducibility of 0.0001 degrees or better with accuracy within 0.01 degrees
• Whether sample stage and detector move together or independently must be stated
• Note: larger radius improves resolution but changes sensitivity and footprint, so match to sample type
3. Detectors
• One dimensional array or two dimensional area detectors
• Energy resolution and maximum count rate should be specified
• Zero background mode or filters to suppress fluorescence
• Note: detector performance decides trace phase detection more than tube power does
4. Angular accuracy and resolution
• Peak position accuracy and full width at half maximum must be measured values, not claims
• Resolution is normally calibrated with standards such as LaB6 or silicon
• Instrument broadening must be controlled, or crystallite size and microstrain results are distorted
• Note: measured width plus standard reference is the acceptance criterion worth writing into the contract
5. Sample stages and attachments
• Powder holders, bulk sample stages and grazing incidence stages as basics
• Optional high temperature stage up to 1600 degrees Celsius, cryogenic stage and in situ cells
• Sample spinner to reduce preferred orientation
• Six or nine position autosamplers for batch work
• Note: in situ attachments add most of the value and should be ordered with the instrument
6. Software and databases
• Phase identification needs a PDF card database such as ICDD with stated coverage and licence
• Quantitative analysis with Rietveld refinement, K value method or internal standard
• Crystallite size and microstrain via Scherrer and Williamson Hall methods
• Raw data export for independent re analysis
• Note: database licences and software modules are often priced separately, so the quote must list them
7. Detection limits and quantification
• Typical phase detection limits around 1 to 3 percent by mass, with longer scans for trace phases
• Quantitative phase accuracy generally 1 to 5 percent absolute
• Standardless quantification requires a stated algorithm and working range
• Note: detection limits depend strongly on scan time, so test conditions must accompany the figures
8. Radiation safety and installation: the key item
• X-ray diffractometers are radiation producing devices, requiring a safety licence and certified operators
• Door interlocks, warning lamps, dose monitoring and emergency stop are mandatory
• Room walls and doors must meet radiation shielding requirements
• Stabilised power, cooling water loop and temperature control within 1 degree Celsius
• Note: the licence is a prerequisite, and sites often discover it is missing after delivery
9. Where procurement goes wrong
• Judging by tube power alone while ignoring detector and optics, so trace phases remain undetectable
• Software modules and PDF databases excluded from the quote, with core functions charged later
• Ignoring licence and room construction lead time, leaving the instrument unable to pass acceptance