Micromanipulator Specifications

Home / Specifications Library / Micromanipulator Specifications

Micromanipulator Specifications

2026-10-27

1. Mechanical structure and travel

• Three axes, X Y Z, or four including diagonal or rotation

• Coarse travel of 20 to 25 mm for rapid approach

• Fine travel of 2 to 5 mm by micrometer screw or piezo drive

• Left or right hand configuration, with two arms mountable together

• Rigidity and backlash directly affect positioning repeatability

2. Positioning precision: the key item

• Fine mode steps of 0.1 micrometre or less

• Repeatability of 1 micrometre or better when returning to a target

• Drift of 1 micrometre per hour or less for prolonged work such as nuclear transfer or ICSI

• Oil free mechanisms without backlash are more stable, while hydraulic types need maintenance against leaks

• Note: precision must be measured under working load with a pipette fitted, since unloaded figures mean little

3. Joystick and proportional control

• Three axis coaxial joystick aligned with the direction of travel to reduce misoperation

• Proportional control where pipette travel follows handle travel, with adjustable sensitivity

• One touch switching between coarse and fine

• Foot switch for injection and holding vacuum

• Adjustable damping, since too light causes tremor and too heavy feels vague

4. Injection and holding systems

• Injection by air or oil pressure, or by piezo drive

• Piezo drive suits hard shelled samples such as nematodes and zebrafish embryos

• Adjustable injection volume down to the picolitre range

• Holding pipette vacuum adjustable through a gas tight syringe or oil system

• Capillary holder sizes must match the pipettes in use

5. Microscope compatibility

• Designed for inverted microscopes, with the brand and model interface confirmed

• Usually mounted on the stage sides or a dedicated side plate

• The stage should be fixed or accept an operating platform so it does not shift during work

• Compatibility with incubation stages and heated stages must be checked in advance

• Working distance of a 40X objective must accommodate the pipette angle

6. Vibration isolation and environment

• An anti-vibration table, air or marble, is mandatory

• Away from lifts, compressors, corridors and doors

• No draughts from air conditioning across the working area

• Stable temperature of 20 to 25 degrees Celsius, since drift follows temperature change

• Level floor and levelled instrument

7. Accessories and consumables

• Pipette puller and microforge as companion equipment

• Consumables including injection and holding pipettes and capillaries, with sizes and cost stated

• Automatic pipette change or position memory on some systems

• Cleaning and dust cover kit

8. Power and environment

• Power supply AC 220 V, 50 Hz

• Operating temperature 20 to 25 degrees Celsius, drift within 1 degree per hour

• Relative humidity 40 to 70 percent

• Class 10000 clean room or laminar bench for cell work

• Independent earth required

9. Where procurement goes wrong

• Only micromanipulator is specified without step size or drift, and ICSI or nuclear transfer fails because drift exceeds the working tolerance

• The interface does not match the existing inverted microscope, requiring an adapter plate or a new stand

• No anti-vibration table and no temperature control, so micrometre work cannot be held steady on an ordinary bench

Need this kind of equipment, or advice on configuration? Tell us your sample types and objectives — we will come back with models and pricing.