Infrared Microscope Specifications

Home / Specifications Library / Infrared Microscope Specifications

Infrared Microscope Specifications

2026-10-27

1. Optical system

• Visible and infrared light paths, the visible for alignment and the infrared for acquisition

• Path change by motor or slider, without moving the sample

• Gold coated or moisture protected optics for humid environments

• Purgeable with dry air or nitrogen to suppress water vapour and CO2 interference

2. Spectral range and resolution: the key item

• Mid infrared range 4000 to 650 reciprocal centimetres as standard, extended models to 400

• Resolution of 4 reciprocal centimetres or better, with 1 to 2 needed for fine structure

• Wavenumber accuracy better than 0.1 reciprocal centimetres, verified on polystyrene

• Indicative signal-to-noise better than 10000 to 1 for a single point at 4 over one minute

• Note: infrared microscopy gives far less signal than a bench FT-IR, so signal-to-noise follows spot size

3. Detectors

• MCT detector, high sensitivity and fast, needing liquid nitrogen, the usual choice for microscopy

• DTGS detector, room temperature without nitrogen, lower sensitivity, suited to large spots

• Cooling method must be stated clearly, nitrogen dewar or electrically cooled, as it drives maintenance

• Nitrogen capacity and refill interval should be written into acceptance terms

4. Objectives and ATR

• Infrared objectives at 15X and 32X, both reflective Cassegrain, never ordinary glass optics

• ATR objectives in diamond, germanium or silicon for pressing onto opaque and thick samples

• Spot size down to 10 micrometres with a 32X ATR, to match the feature of interest

• Transmission mode needs KBr or ZnSe windows and samples under 10 micrometres thick

• Reflection mode suits coatings, films and surface contamination

5. Mapping stage

• Motorised X Y stage with at least 75 by 50 mm travel

• Minimum step of 1 micrometre or less for point by point mapping

• Chemical imaging from a grid of spectra, giving component distribution and peak intensity maps

• Line and area scans with predictable point counts and acquisition time

• Stage coordinated with visible imaging so the sample is seen before it is measured

6. Sample viewing and alignment

• Colour camera on the visible path for live viewing and point selection

• Crosshair or marker overlay so the measured point matches the visible location

• Spot position verified on a standard sample against the crosshair

• Large samples limited by stage travel and objective working distance

7. Software and libraries

• Acquisition with baseline correction, smoothing and normalisation

• Library search against commercial and user libraries, including mixtures

• Mapping processing with chemical imaging, principal component analysis and clustering

• Export to JCAMP-DX, CSV and TIFF

• Compliance features such as audit trail and electronic signature as options

8. Power and environment

• Power supply AC 220 V, 50 Hz

• Operating temperature 15 to 30 degrees Celsius, drift within 2 degrees per hour

• Relative humidity up to 60 percent, lower for KBr windows

• Away from vibration, since mapping is sensitive to it

• Safe storage and handling of liquid nitrogen must be arranged in advance

9. Where procurement goes wrong

• Only infrared microscope is specified without detector type or spot size, and a 100 micrometre spot arrives, making micro analysis impossible

• An MCT detector is supplied without a nitrogen routine, so it stays warm and loses sensitivity

• No ATR objective, so opaque and thick samples cannot be measured without destructive preparation