Ultra-Depth-of-Field 3D Microscope Specifications

Home / Specifications Library / Ultra-Depth-of-Field 3D Microscope Specifications

Ultra-Depth-of-Field 3D Microscope Specifications

2026-10-27

1. Optics and magnification range

• Motorized optical zoom lens, zoom range 0.35X to 5.0X as an indication, depending on the model

• Optical magnification continuously adjustable from 20X to 2000X, with digital magnification extending the displayed value

• Optical design with large depth of field, several times that of a conventional microscope in a single pass

• Optical and digital magnification must be stated separately, displayed magnification is not optical magnification

2. Depth-of-field stacking: the core function

• Motorized Z scanning, capturing layer by layer and combining into an all-in-focus image

• Z travel of 30 mm or more, up to 100 mm on long-travel models

• Number of layers per stack configurable, typically 20 to 200

• Three outputs: all-in-focus image, 3D surface reconstruction and height map

• Note: stacking is the value of the system, not simply enlarging the picture

3. Resolution and pixels

• Camera of 5 megapixels or more, mainstream 12 to 20 megapixels

• Sensor size of 1/1.8 inch or larger, larger sensors gathering more light per pixel

• Best optical resolution of 1 micrometre or finer depending on model and objective

• Pixel resolution changes with zoom and must be calibrated in software

4. XYZ travel and stage

• Motorized X and Y stage, travel of at least 50 x 50 mm, supporting image stitching

• Motorized Z with step resolution of 1 micrometre or finer

• Load capacity of 3 kg or more, taking parts, circuit boards and workpieces

• Mosaic stitching for very large fields

5. Measurement functions

• 2D measurement of distance, angle, radius, area and parallelism

• 3D measurement of height difference, volume, roughness Ra and Rz, and flatness

• Calibration with a standard gauge block before measuring, calibration accuracy sets the credibility of the data

• Measurement reports exportable with images, data tables and pass or fail results

6. Illumination

• Multi-segment ring illumination, lit in sections to reveal surface relief and scratches

• Coaxial illumination for mirrors and highly reflective samples

• Oblique illumination for edges, burrs and recesses

• Transmitted illumination as an option for transparent samples

• Brightness and angle controlled in software, allowing comparison across lighting conditions

7. Software and data

• Live depth stacking, 3D display and rotational viewing

• Batch measurement and templated inspection routines

• Images in JPEG, TIFF or BMP, 3D data exported in common formats

• Customisable report templates with company branding

8. Power and environment

• Power supply AC 220 V, 50 Hz

• Operating temperature 5 to 40 degrees Celsius, relative humidity up to 80 percent

• Anti-vibration table recommended, vibration affects high magnification and 3D measurement noticeably

• Precision optics plus motorized stage, so regular calibration is required

9. Three points where procurement often goes wrong

• Maximum displayed magnification is taken for optical magnification, where optical is only a few tens of times and enlargement turns to mosaic

• Z step accuracy or stage repeatability is inadequate, giving height errors above 10 percent

• No gauge block supplied and no calibration routine, so measurements cannot be traced and reports cannot support a verdict

Need this kind of equipment, or advice on configuration? Tell us your sample types and objectives — we will come back with models and pricing.
☎Call us ✉Enquiry
13838311486 Same number on WeChat